During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.
Source facts
Action
There are two field safety notices, one for Consignees who have surgeons that use the impaction fixtures (Verson 1) and one for Consignees who had previously indicated their surgeons do NOT use the impaction fixtures (Verson 2). The two recall notification letters were sent out on 2/9/17.
Initiated
2017-02-09
Root cause
Device Design
Terminated
2017-10-20
Distribution
US, South Korea, Australia, Canada, United Kingdom/Ireland, Germany
Recall number
Z-1412-2017
Recall status
Terminated
CDRH record ID
153244
FDA record created
2017-03-08
Recall event number
76456