Open original FDA record ↗Recall record
- FDA identifier
- Z-1412-2019
- Company in FDA record
- Elekta Instrument AB
- FDA record creation date
- May 23, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The margin tool in Leksell GammaPlan 11.1 systematically overestimates margin in certain areas of the volume.
Source facts
- Action
- Elekta initially notified customers on about 06/27/2018 via "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" LETTER. Instructions included to post the recall notice in a place accessible to all users, e.g. Instructions for Use, until the recall action is closed, advise the appropriate personnel working with the affected device of the recall and the content of the recall notification letter, and complete and return the Acknowledgement Form. Elekta updated their recall strategy on about 11/21/2018 and will be providing all affected customers with software upgrade 11.1.1. Electa service engineers will be pro-actively reaching out to customers to organize a date for installation on all affected devices. An updated recall notification was not sent to customers.
- Initiated
- 2018-06-27
- Root cause
- Software design
- Distribution
- Worldwide Distribution - US Nationwide distribution to GA, MD, NJ, PA, PR, WA. International distribution to Australia, Belgium, France, Germany, Hong Kong, Italy, Japan, Korea, Democratic People's Republic of, Netherlands, Romania, Russian Federation, Spain, Turkey, United Kingdom.
- Recall number
- Z-1412-2019
- Recall status
- Open, Classified
- CDRH record ID
- 172309
- FDA record created
- 2019-05-23
- Recall event number
- 82662