Z-1413-2021: cobas SARS-CoV-2 & Influenza A/B assay cobas Liat- Reagents, 2019-Novel Coronavirus Nucleic Acid for use

Recall record Z-1413-2021 for Roche Molecular Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1413-2021
Company in FDA record
Roche Molecular Systems, Inc.
FDA record creation date
Apr 13, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Invalid or false positive results for one or more targets (SARS-CoV-2/Influenza A/Influenza B) using the cobas SARS-CoV-2 & Influenza A/B assay for use under emergency use authorization on the cobas Liat analyzer, showed abnormal PCR growth curves

Source facts

Action
Roche issued Urgent Medical Device Correction letter on March 25, 2021. Letter states reason for recall, health risk and action to take: Actions Required " Inform any operators that use the cobas Liat analyzer and/or cobas SARS-CoV-2 & Influenza A/B test of the potential hazards associated with this issue and provide a copy of this UMDC as appropriate. " If you observe an increase in Influenza B positive results that does not match community infection rates, stop testing on the cobas Liat analyzer and contact the Roche Support Network Customer Support Center at 1-800-800-5973 to determine if a tube leak or other issue has occurred. " If you observe dual- and/or triple-positive results for a particular sample, retest the sample with the cobas SARS-CoV-2 & Influenza A/B test for use with the cobas Liat analyzer using a different cobas Liat analyzer. If you do not have an additional analyzer or if results on the second analyzer disagree with the results from the first analyzer, stop testing on the affected analyzer and contact the Roche Support Network Customer Support Center at 1-800-800-5973. " Monitor your invalid result rates. If there is a sudden increase or if invalid results are observed repeatedly on a specific cobas Liat analyzer, stop testing on the analyzer and contact the Roche Support Network Customer Support Center at 1-800-800-5973. " If the Thermal runaway (disable) message appears (as described later in this UMDC) on your cobas Liat analyzer running software version 3.3.0, stop testing on the analyzer and contact the Roche Support Network Customer Support Center at 1-800-800-5973. " For handling previous positive results generated with the cobas SARS-CoV-2 & Influenza A/B test for use on the cobas Liat analyzer, follow your laboratorys standard operating procedures to investigate the potential for false positive results. " Complete all sections of the enclosed faxback form (TP-01256) and fax it to 1-866-503-1216 or email it to roche6136@ster
Initiated
2021-03-25
Root cause
Unknown/Undetermined by firm
Distribution
Nationwide WorldWide
Recall number
Z-1413-2021
Recall status
Open, Classified
CDRH record ID
186412
FDA record created
2021-04-13
Recall event number
87622

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗