Z-1415-2023: BD Kiestra InoqulA+ with BD Kiestra" InoqulA+" and BeA Software version 5.1 and 5.1.1 as part of the Icef

Recall record Z-1415-2023 for Becton Dickinson & Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1415-2023
Company in FDA record
Becton Dickinson & Co.
FDA classification date
Apr 20, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Upon installation of BD Kiestra" InoqulA+" and BeA version 5.1 and 5.1.1 as part of the Icefall A platform, plate information was not visible in Synapsys" after processing has occurred. This could possibly lead to a delayed diagnosis and/or treatment or need for specimen recollection streaking plated media, and inoculating tubes and slides. In Semi-Automated (SA) mode, plates are automatically selected, barcoded, streaked in a pre-configured pattern while the user manually inoculates plates, tubes, and slides. An optional biosafety cabinet on the SA module provides personnel, product, and environmental protection. The InoqulA+" solution is indicated for use in the clinical laboratory.

Source facts

Action
BD issued URGENT: Medical Device Product Correction on 3/22/23 via email. Letter states reason for recall,, health risk and action to take: 1) Read and understand the content of this letter. 2) Share and post this recall letter within your facility network to ensure awareness. 3) Customers should utilize the module of the Icefall platform identified by their regional BD service team that does not exhibit this issue to avoid possible unavailability of results. 4) Complete the attached Customer Response Form and return to the BD contact noted so that BD may acknowledge your receipt of this notification per FDA requirements. Taken by BD: 1) BD has developed and remotely implemented an interim solution to immediately address the software issue and prevent disruption of testing for the patients. 2) BD is taking further action to prevent recurrence of this product issue. Contact Information: US Contact Information Areas of Support North American Regional Complaint Center, Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Recall when prompted. Mon Fri 8:00am and 5:00pm CT or Email: productcomplaints@bd.com
Initiated
2023-03-22
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution in the states of IL, NY and the countries of Netherlands, Saudi Arabia.
Recall number
Z-1415-2023
Recall status
Open, Classified
CDRH record ID
199239
FDA classified
2023-04-20
Recall event number
91937

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗