Skip to content

Public FDA record

Z-1416-2021: BioGX SARS-CoV-2 Reagents for BD MAX System-IVD PCR test intended for the qualitative detection of nucle

Recall record Z-1416-2021 for Becton Dickinson & Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1416-2021
Company in FDA record
Becton Dickinson & Co.
FDA record creation date
Apr 13, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

BioGX SARS-CoV-2 Reagents for BD MAX" System rehydration buffer tubes were not adequately sealed causing some tubes to be partially filled or empty upon use. A partially filled or empty rehydration buffer tube may lead to an Unresolved (UNR) result on the BD MAX" System

Source facts

Action
BD issued Urgent Medical Device Correction letter on March 24, 2021 FedEx tracked. Letter states reason for recall, health risk and action to take: 1. Inspect rehydration buffer tubes prior to use for the specific catalog and lot numbers listed below in Attachment A and destroy any products containing partially filled or empty tubes following your institutions process for destruction. 2. Share and Post this medical device recall notification within your facility network to ensure users inspect tubes prior to use. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your product replacement, if required. 4. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Actions Taken by BD: 1. Corrective actions have been initiated to prevent recurrence of the identified root cause. 2. BD will provide a product replacement for all destroyed inventory. If you require further assistance, please contact: BD Contact US Contact Information Outside of US BD Customer/ Technical Support 800-638-8663 Monday Friday 7:00am and 7:00pm EST For customers outside the US, contact the local BD representative or distributor.
Initiated
2021-03-24
Root cause
Under Investigation by firm
Terminated
2022-08-02
Distribution
US Nationwide
Recall number
Z-1416-2021
Recall status
Terminated
CDRH record ID
186392
FDA record created
2021-04-13
Recall event number
87605

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗