Skip to content

Public FDA record

Z-1417-2023: 2D Perfusion, release R1.0.x, R1.1.x, R1.2, and R1.2.1

Recall record Z-1417-2023 for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1417-2023
Company in FDA record
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
FDA classification date
Apr 20, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There are technical issues related to signal generation and processing, which can lead to inaccurate presentations.

Source facts

Action
A recall notification letter was issued February 20, 2023. The Urgent Medical Device Correction letter includes the following instructions for the Customer: " Stop using SmartPerfusion and 2D Perfusion. NOTE: This message does not impact (other parts of) the Philips Allura or Azurion systems or other tools within the Interventional Workspot. The Philips Allura and Azurion systems as well as other Interventional Workspot tools may continue to be used. " Place the Urgent Medical Device Correction Letter with the documentation of the system. " Circulate the notice to all users so that they are aware of the product issue. " Return the reply form attached to the Urgent Medical Device Correction letter to Philips to confirm that the users of the system have reviewed and understood it.
Initiated
2023-02-20
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution.
Recall number
Z-1417-2023
Recall status
Open, Classified
CDRH record ID
198921
FDA classified
2023-04-20
Recall event number
91806

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.