Potential for health care provider/hospital staff member to sustain injury to fingers or hands when removing/handling the headrest component of the VERTIER Surgical Table.
Source facts
Action
An "URGENT: Device Recall" letter dated May 11, 2009 was issued to consignees. Consignees were informed that Stryker service representatives will visit each location to affix a caution label identifying the pinch point on each headrest. Additionally, consignees will be provided with an updated surgical table user manual. Direct questions to Stryker Communications Corporation by calling 972-410-7100.
Initiated
2009-05-08
Root cause
Device Design
Terminated
2010-04-26
Distribution
Nationwide Distribution and Canada.
Recall number
Z-1418-2009
Recall status
Terminated
CDRH record ID
81848
FDA classified
2009-05-26
Recall event number
51994