- FDA identifier
- Z-1419-2021
- Company in FDA record
- Covidien Llc
- FDA record creation date
- Apr 13, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for the pouch to contain an extra laminate layer, creating the potential for the device sterility to be compromised. The inclusion of an additional laminate layer in the package may also impact complete de-gassing of the sterilization gas, Ethylene Oxide and may result in a potentially increased risk for patient infection and/or potential exposure to Ethylene Oxide retained in the pouch
Source facts
- Action
- Medtronic notified consignees by letter delivered via Federal Express and/or certified mail beginning March 23, 2021. The letter states reason for recall, health risk and action to take: " Immediately quarantine and discontinue use of the affected item code and lot. " Return affected product. " Forward the information to anyone who they have distributed affected Covidien Tri-Staple" 2.0 Reinforced Intelligent Reloads. " Complete the Recalled Product Return Form Questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
- Initiated
- 2021-03-23
- Root cause
- Nonconforming Material/Component
- Terminated
- 2024-05-15
- Distribution
- AL, AR, FL, IN, LA, MA , MN, MO, NJ, NY, OR, PA, TX
- Recall number
- Z-1419-2021
- Recall status
- Terminated
- CDRH record ID
- 186407
- FDA record created
- 2021-04-13
- Recall event number
- 87618