Open original FDA record ↗Recall record
- FDA identifier
- Z-1419-2022
- Company in FDA record
- DePuy Orthopaedics, Inc.
- FDA record creation date
- Jul 20, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Manufactured with a Galvanized High Carbon Steel spring instead of the correct 302 Stainless steel spring. has the potential to oxidize and corrode in the ATTUNE Measured Sizing & Rotation Guide, resulting in surgical delay, debris in surgical field, Osteolysis, Poor Joint Mechanics, Soft Tissue Irritation, Pain
Source facts
- Action
- DePuy Synthes issued Urgent Medical Device Correction Letter on June 29 2022. Letter states reason for recall, health risk and action to take: 1. Examine your inventory immediately for the subject lots and quarantine the products. 2. To minimize disruption, replacement of the spring will be completed by a Field Service Representative. You will be contacted directly by a member of the Field Service Team. 3. Familiarize yourself with the content of this letter and forward this notice to any personnel in your facility who need to be informed. 4. Complete all fields, sign, and return the attached business response form (BRF) on the last page of this letter to OneMD-Field-Actions@its.jnj.com within three (3) business days of receipt of this medical device correction letter. Please include FA 2126657 - Attune Measured Sizing & Rotation Guide in the email subject line. 5. Complete the BRF even if you do not have any of the subject product in your current inventory. 6. If any of the subject product has been forwarded to another facility, contact that facility, and provide a copy of this notice to the relevant personnel. 7. Post a copy of this notice in a visible area for awareness of this correction. Contact your local DePuy Synthes Sales Consultant. For Medical Information Requests, please visit our website: https://www.jnjmedicaldevices.com/mir
- Initiated
- 2022-06-29
- Root cause
- Process control
- Distribution
- Nationwide Foreign: Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Lithuania, Mauritius, Mexico, Morocco, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Vietnam
- Recall number
- Z-1419-2022
- Recall status
- Open, Classified
- CDRH record ID
- 194778
- FDA record created
- 2022-07-20
- Recall event number
- 90557