Z-1419-2023: Beckman Coulter Access Free T3, Triiodothyronine, Free, Catalog #A13422, each box contains 2 reagent pack

Recall record Z-1419-2023 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1419-2023
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Apr 20, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The reagent lot produces elevated results (approximately >30%) in some (15-20%)patient samples that do not correlate with the results of other thyroid markers nor Free T3 results generated on other platforms. This introduces a risk of the positive bias not being detected by QC with the potential that clinical correlation with other thyroid function tests will be inconsistent requiring additional testing; however the clinical assessment and treatment decisions are not made in isolation on a single test result especially, in the setting of thyroid function testing. False positive test results could lead to a euthyroid patient being interpreted clinically as having elevated thyroid function (hyperthyroid) or a hypothyroid patient being within the normal range. Such misdiagnoses may lead to inappropriate treatment of a euthyroid patient with anti-thyroid medications or hypothyroid patient not getting sufficient treatment since their free T3 test result demonstrates them to be in the euthyroid range.

Source facts

Action
The recalling firm issued letters dated 3/8/2023 via mail and email. The letter described the issue, impact, and actions to be taken. The actions included discontinuing use of the affected lot and discard all remaining reagent packs from this lot. The letter is to be shared with the consignee laboratory and/or medical director regarding the need to review previous patient test results. The consignee is to contact their local Beckman Coulter representative for replacement product requests. If the product has been forwarded to another laboratory, the consignee is requested to provide them a copy of the letter. A response was requested within 10 days via email or completion and return of the enclosed response form. The response form is to indicate the letter has been read and understood and all relevant personnel have been informed of its contents. The consignee is also to indicate whether or not they had the affected product.
Initiated
2023-03-15
Root cause
Process design
Distribution
Worldwide distribution - US Nationwide, including Puerto Rico and the countries of Bahrain, Brazil, Canada, China, Costa Rica, El Salvador, Georgia, Hong Kong, Malaysia, Mexico, Myanmar, New Zealand, Peru, Philippines, Qatar, Singapore, Taiwan, Thailand, Uruguay, and Viet Nam.
Recall number
Z-1419-2023
Recall status
Open, Classified
CDRH record ID
199210
FDA classified
2023-04-20
Recall event number
91916

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗