Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.
Source facts
Action
Autochair will issue a Retrofit Failsafe Kit. In the meantime, please quarantine current stock of SmartBase products until the failsafe device has been fitted and tested.
Initiated
2018-03-07
Root cause
Component design/selection
Distribution
PA, and worldwide
Recall number
Z-1420-2018
Recall status
Open, Classified
CDRH record ID
162854
FDA record created
2018-04-18
Recall event number
79627
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.