Z-1420-2020: AIA-360 Automated Immunoassay Analyzer, Product Code 019945

Recall record Z-1420-2020 for Tosoh Bioscience Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1420-2020
Company in FDA record
Tosoh Bioscience Inc
FDA classification date
Feb 11, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A display screen software issue on the analyzer causes the display to freeze when display screen is touched at the same time as a command from the instrument firmware, causing the instrument to stop. As a result, the run is aborted and the results are not retrievable.

Source facts

Action
The firm distributed Urgent Medical Device Correction letters to customers on January 10, 2020. Customers were instructed to retrain from touching the display panel while the analyzer is running. If the analyzer stops and/or the display freezes, please contact Tosoh Technical Support 24 hours a day, seven days a week at (800) 248-6764. If your analyzer's serial number is listed among the recalled units, you will be contacted by a Tosoh Field Service Engineer to schedule a visit to replace the display screen which will correct the potential issue. Should you have any questions regarding this medical device recall, please feel free to contact Bernadette OConnell at (800) 248-6764 or by email at bernadette.oconnell@tosoh.com. She will be available to answer any questions Monday - Friday, from 9:00 AM to 5:00 PM (PST).
Initiated
2020-01-10
Root cause
Software design
Terminated
2020-09-10
Distribution
Distributed to AZ, NE, NJ, NY, and WA. Foreign distribution to Cayman Islands, Costa Rica, and Uruguay
Recall number
Z-1420-2020
Recall status
Terminated
CDRH record ID
179409
FDA classified
2020-02-11
Recall event number
84787

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗