Open original FDA record ↗Recall record
- FDA identifier
- Z-1421-2009
- Company in FDA record
- Neuronetics, Inc.
- FDA classification date
- May 27, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
SenStar Treatment Link may have an open circuit that during use with the NeuroStar TMS Therapy system may cause arcing and heat damage to the SenStar and/or treatment coil. Heating of the SenStar during use has a potential risk of minor scalp burn to patients.
Source facts
- Action
- Neuronetics Inc. notified consignees by telephone and sent a follow-up fax/letter dated March 19, 2009 titled "Urgent Product Recall Notification." The letter describes the issue and actions for consumers to take in response to the recall. Neuronetics will visit each customer in person to complete a "Returned Goods: SenStar Recall Inventory Form." Customer questions should be directed to Neuronetics Customer Service at 1-877-600-7555.
- Initiated
- 2009-03-16
- Root cause
- Process control
- Terminated
- 2009-05-29
- Distribution
- Nationwide Distribution to AZ, CA, DC, GA, FL, IL, MD, MI, MN, NH, NY, PA, RI, SC, and WA.
- Recall number
- Z-1421-2009
- Recall status
- Terminated
- CDRH record ID
- 81035
- FDA classified
- 2009-05-27
- Recall event number
- 51703