Open original FDA record ↗Recall record
- FDA identifier
- Z-1421-2011
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Feb 23, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Firm became aware of a potential malfunction and possible hazard to patients in the event the tilt buttons on the OPTI Grip become stuck under the housing
Source facts
- Action
- Siemens Medical Solutions USA, Inc issued a Customer Safety Advisory Notice via Update Instruction XP005/10/S to affected customers. This letter informs customers of the potential malfunction and of possible danger for patients, users or other persons. A future Update Instruction will be released for the replacement of the tilt buttons on all OPTI Grip handles. Customers were instructed to forward the safety notice to any new owners of this device and to provide Siemens Medical with names of the new customers. For any questions regarding this recall call 610-448-3237.
- Initiated
- 2011-01-14
- Root cause
- Other
- Terminated
- 2013-07-16
- Distribution
- Nationwide Distribution including AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY.
- Recall number
- Z-1421-2011
- Recall status
- Terminated
- CDRH record ID
- 97376
- FDA classified
- 2011-02-23
- Recall event number
- 57775