Alphatec Spine initiated this medical device recall for the following reason: due to tolerancing, the flange of the Phygen Leucadia" Autolok set screw may interfere with the top of the screw head prior to fully locking the rod. If this condition occurs, the polyaxial
screw head and rod may fail to provide a rigid construct under loading. To date there have been
no reports of patient injury, adve
Source facts
Action
Alphatec Spine initiated recall by contacting customers via telephone and/or email on 05/06/2013. An Urgent Medical Device Recall Notification letter along with a response form were sent via FedEx to all affected accounts. The letter identified the product, problem, and actions to be taken by the customers.
Initiated
2013-05-06
Root cause
Device Design
Terminated
2014-09-25
Distribution
Nationwide Distribution.
Recall number
Z-1421-2013
Recall status
Terminated
CDRH record ID
118068
FDA classified
2013-05-28
Recall event number
65165