Open original FDA record ↗Recall record
- FDA identifier
- Z-1422-2009
- Company in FDA record
- St. Jude Medical
- FDA classification date
- Jun 12, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Non-compliance with IEC Standard 60601-1 Clause 57.6 relative to fusing of AC input line. The neutral line is not fused. The hot line in fused. Clause 57.6 calls for both lines to be fused.
Source facts
- Action
- All foreign customers were sent Urgent Product Safety Information letters on March 23, 2009. Letters described units affected, the reason for action, and the actions to take to eliminate risk. It is recommended that the information in the letter be circulated to all users, risk managers and the biomedical engineering department responsible for maintaining and inspecting electrical equipment located within their institution. The letter also stated that a SJM representative will be in touch to perform the EP-4 Stimulator fuse modification. Customers are to complete and return the attachment.
- Initiated
- 2009-03-23
- Root cause
- Device Design
- Terminated
- 2010-01-20
- Distribution
- Worldwide distribution.
- Recall number
- Z-1422-2009
- Recall status
- Terminated
- CDRH record ID
- 80957
- FDA classified
- 2009-06-12
- Recall event number
- 51681