The firm become aware of a potential issue with ARTISTE system that may result in patient injury should the positioner belt break, and the cam disc (secondary safety mechanism) is not installed and if the flat panel is over the patient.
Source facts
Action
The firm, Siemens, issued a "Customer Information" letter on December 3, 2010 via mail to all its direct consignees/customers. The letter described the product, problem and actions taken by the firm to the customer. Siemens arranged for an inspection of their customers systems to make sure the safety mechanism is fully functional. Siemens instructed the customers to include the Customer Information letter in their ARTISTE System Owner manual chapter 'Safety Advisory Letters' .
If you any questions, please call 925-246-8407.
Initiated
2010-12-03
Root cause
Other
Terminated
2011-06-30
Distribution
Worldwide distribution.
Recall number
Z-1422-2011
Recall status
Terminated
CDRH record ID
97050
FDA classified
2011-02-24
Recall event number
57704