Skip to content

Public FDA record

Z-1424-2009: Medtronic, Transfer Spike, Blood Processing Accessory, REF BT945, Sequestration Kits

Recall record Z-1424-2009 for Medtronic Cardiovascular Revascularization & Surgical Therap. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1424-2009
Company in FDA record
Medtronic Cardiovascular Revascularization & Surgical Therap
FDA classification date
Jun 23, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Sterility may be compromised. Non-vented caps that connect to the 3-way stopcock attached to the transfer spike could prevent Ethylene Oxide from contacting the inside surfaces of the stopcock during sterilization, leading to a potentially non-sterile product. Use of the product may cause serious health consequences from potential sepsis/infection and the associated remote risk of death.

Source facts

Action
Consignees were sent a Medtronic "Urgent Medical Device Recall Notice" letter dated March 5, 2009. The letter was addressed to Risk Manager. The letter described the product and problem and requested the consignee to complete, sign and return the certificate acknowledgement.
Initiated
2009-02-25
Root cause
Device Design
Terminated
2011-05-07
Distribution
Distribution to one medical facility in MN.
Recall number
Z-1424-2009
Recall status
Terminated
CDRH record ID
81730
FDA classified
2009-06-23
Recall event number
51952

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗