Recall record Z-1424-2009 for Medtronic Cardiovascular Revascularization & Surgical Therap. View source facts, labeled dates, and the original FDA record.
Sterility may be compromised. Non-vented caps that connect to the 3-way stopcock attached to the transfer spike could prevent Ethylene Oxide from contacting the inside surfaces of the stopcock during sterilization, leading to a potentially non-sterile product. Use of the product may cause serious health consequences from potential sepsis/infection and the associated remote risk of death.
Source facts
Action
Consignees were sent a Medtronic "Urgent Medical Device Recall Notice" letter dated March 5, 2009. The letter was addressed to Risk Manager. The letter described the product and problem and requested the consignee to complete, sign and return the certificate acknowledgement.
Initiated
2009-02-25
Root cause
Device Design
Terminated
2011-05-07
Distribution
Distribution to one medical facility in MN.
Recall number
Z-1424-2009
Recall status
Terminated
CDRH record ID
81730
FDA classified
2009-06-23
Recall event number
51952