Z-1425-2012: Clinical Chemistry Total Bilirubin, 6L45-20 and 6L45-40. Used for the quantitative analysis of total b

Recall record Z-1425-2012 for Abbott Laboratories Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1425-2012
Company in FDA record
Abbott Laboratories Inc.
FDA classification date
Apr 16, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Abbott Laboratories has issued a recall on the Clinical Chemistry Total Bilirubin due to an under-recovery on one clinical neonatal specimen resulting in a value less than the assay linear limit.

Source facts

Action
Abbott Laboratories plans to release a Product Correction letter to customers instructing customers to follow the steps provided in order to revise their Total Bilirubin assay parameters. The letter explained the reason for the recall, along with the patient impact. Customers were to keep the letter as part of their laboratory records and also forward a copy to any other laboratories that they may have forwarded Total Bilirubin kits. U.S. Customers should contact Customer Service at 1-877-4ABBOTT, if they have questions.
Initiated
2009-07-09
Root cause
Material/Component Contamination
Terminated
2014-03-07
Distribution
Worldwide Distribution -- USA, Guatemala, Chile, Brazil, Canada, Germany, Hong Kong, Thailand, Puerto Rico, Australia, New Zealand, Trinidad & Tobago, Bermuda, Curacao, Cayman Island, and Jamaica.
Recall number
Z-1425-2012
Recall status
Terminated
CDRH record ID
83856
FDA classified
2012-04-16
Recall event number
52748

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗