Z-1425-2021: Remel Haemophilus Test Medium(100mm) 10/PK, REF: R01503 Remel Haemophilus Test Medium (HTM) is a solid

Recall record Z-1425-2021 for Remel Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1425-2021
Company in FDA record
Remel Inc
FDA record creation date
Apr 13, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When tested with H. influenza ATCC 49247 and ATCC 49766 both organisms may exhibit lighter than normal growth making zones difficult to measure.

Source facts

Action
The firm, ThermoFisher Scientific, sent an "Medical Device Field Action Recall" letter dated 3/10/2021 to its consignees on 03/10/2021. The letter described the product, problem and actions to be taken. The consignees were instructed to do the following: Review the reported test results should be determined by the appropriate technical expert. Please keep this notification on file. This notice needs to be passed on to all who need to be aware within your organization or to any organization where the potentially affected products have been transferred. If you have any questions, please contact Technical Services Department at 800-255-6730 (US) or email microbiology.ts.us@thermofisher.com. You should complete the accompany Acknowledgment Form upon receipt and return to MBD.vigilance@thermofisher.com. If you have any further questions, contact Director, Quality Assurance and Regulatory Affairs at 913-895-4077 or email: gary.klaassen@thermofisher.com Copy to: Beverly Marsh - beverly.marsh@thermofisher.com.
Initiated
2021-03-10
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-1425-2021
Recall status
Open, Classified
CDRH record ID
186187
FDA record created
2021-04-13
Recall event number
87495

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗