St. Jude Medical received two complaints from customers involving the Telesheath Left Atrial Introducer System side arm detaching from the hub of the inner sheath. Subsequent testing of Telesheath inventory at St. Jude Medical indicated that the side arm from the inner sheath can detach during routine handling and manipulation, and occurs on a random basis.
Source facts
Action
100% of customers were notified by a personal visit by a St. Jude Medical representative, starting 8/4/2004. Letters were given by the representative to the consignee concerning the recall.
Initiated
2004-08-04
Root cause
Other
Terminated
2005-11-21
Distribution
US, Canada, Belgium, Hong Kong, and Japan.
Recall number
Z-1426-04
Recall status
Terminated
CDRH record ID
34648
FDA classified
2004-09-01
Recall event number
29848