Z-1426-2021: Instructions for Use and Patient Manual for HeartWare HVAD System. Accompanies these system components: H

Recall record Z-1426-2021 for Heartware, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1426-2021
Company in FDA record
Heartware, Inc.
FDA record creation date
Apr 16, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Medtronic HeartWare HVAD System Instructions for Use (IFU) and Patient Manual (PM) updates regarding: Carrying Cases, Driveline Cover Orientation, and Controller Power Up Sequence issues. The PM is being updated to clarify use of the support strap and wear instructions in addition to cleaning instructions of the carrying cases. The IFU and PM are being updated to add a useful life for the carrying cases. The PM is being updated to inform users to keep the driveline cover on when disconnecting and reconnecting the driveline. The IFU and PM are being updated to clarify the power-up sequence that causes the alarm indicator LEDs and both sets of battery LEDs to turn red for 2.5 seconds while the LCD displays the power-on message. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87603 and 87604

Source facts

Action
The firm, Medtronic, sent an "Urgent Medical Device Notice" dated Feb-2021 beginning 26-Feb-2021, UPS 2-day delivery, to its consignees who Medtronic records indicate they have received affected product. The notice described the product, problem and actions to be taken. The consignees were instructed to do the following: " Please review the updated IFU and PM steps as included in Appendix A.2 and share with patients as needed. " This notice must be shared with all those who need to be aware within your organization or to any organization where potentially affected patients have been transferred. " Please complete the enclosed Customer Confirmation Form and email to RS.CFQFCA@medtronic.com. If you have questions regarding this material, please contact your Medtronic Field Representative or Sr. Customer Quality Director, at 763-526-8583, email: rs.cvgrecalls@medtronic.com or Customer Quality at megan.e.ties@medtronic.com.
Initiated
2021-02-26
Root cause
Device Design
Distribution
Worldwide distribution: US (nationwide) and countries of: ARGENTINA, Armenia, Australia, AUSTRIA, Bahrain, BELGIUM, Brazil, CANADA, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, FINLAND, France, GERMANY, Greece, Hong Kong, Hungary, INDIA, Israel, Italy, Japan, Kazakhstan, Korea, Kuwait, Latvia, LEBANON, LITHUANIA, Luxembourg, Macedonia, Malaysia, NETHERLANDS, NEW ZEALAND, Norway, POLAND, Qatar, Romania, SAUDI ARABIA, Serbia, Singapore, SLOVAKIA, SOUTH AFRICA Spain, SWEDEN, SWITZERLAND, Taiwan, TURKEY, Ukraine, United Arab, Emirates, United Kingdom, and VIETNAM.
Recall number
Z-1426-2021
Recall status
Open, Classified
CDRH record ID
186062
FDA record created
2021-04-16
Recall event number
87456

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗