Reason stated in the source
Sensis/Sensis Vibe SoftwareVD12A, the ComboBox may encounter a disconnect of its communication during the first patient examination of the day or after a longer period of inactivity to the Sensis Vibe system resulting in no vital signs being available.
Source facts
- Action
- Siemens issued a Customer Safety Advisory Notice (AX016/22/S) dated 6/15/22 distributed to customers. Letter states reason for recall, health risk and action to take. The BIOS configuration of the ComboBox in the affected systems will be updated to correct the issue. This corrective action will be distributed via Update Instruction AX015/22/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein
- Initiated
- 2022-06-15
- Root cause
- Device Design
- Distribution
- US Nationwide Distribution AZ FL KY MI OH PR TX
- Recall number
- Z-1426-2022
- Recall status
- Open, Classified
- CDRH record ID
- 194604
- FDA record created
- 2022-07-20
- Recall event number
- 90504