Skip to content

Public FDA record

Z-1426-2023: JULIET Ti LL, LATERAL LUMBAR TI-LIFE PLATE SMALL, REF: JLT-PL 02 08-S, JLT-PL 02 10-S, JLT-PL 02 12-S, JL

Recall record Z-1426-2023 for SPINEART SA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1426-2023
Company in FDA record
SPINEART SA
FDA classification date
Apr 24, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

JULIET Lateral Lumbar Ti-Life Plate, a component of the JULIET Ti Lateral Lumbar Cage, may disassemble during the preparation stage before implantation, or during surgery when removing the implant holder, which could cause surgical delay.

Source facts

Action
On 3/16/23, recall notices were distributed to customers who were told the following: Plate replacement: If the issue occurs, the surgeon should proceed as follows: 1- Unscrew the Plate screw of the cage with the screwdriver, 2- Scrap the current plate assembly, 3- Use a new set of pre-assembled plate/screw (a minimum of 2 sets of pre-assembled plates/screws per size are part of the device listing). The cages are to be used with the corresponding plate or higher plate which means that a plate with the same height than the cage can be used but also a plate with higher height. 4- In the case of 0-degree and 8-degree cages, a pre-assembled plate/screw is not mandatory as described in the surgical technique Acknowledge receipt of this letter by emailing to orders@spineart.us Discontinue use and return any product on hand. Customers with questions can contact firm at 949-313-6400 Monday through Friday, 8:00am to 5:00pm, Pacific, or email at orders@spineart.us
Initiated
2023-03-16
Root cause
Process change control
Distribution
US Nationwide distribution in the states of NH, TX, CA, VA, IL, RI, NC, AZ.
Recall number
Z-1426-2023
Recall status
Open, Classified
CDRH record ID
199335
FDA classified
2023-04-24
Recall event number
91973

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.