Open original FDA record ↗Recall record
- FDA identifier
- Z-1426-2023
- Company in FDA record
- SPINEART SA
- FDA classification date
- Apr 24, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
JULIET Lateral Lumbar Ti-Life Plate, a component of the JULIET Ti Lateral Lumbar Cage, may disassemble during the preparation stage before implantation, or during surgery when removing the implant holder, which could cause surgical delay.
Source facts
- Action
- On 3/16/23, recall notices were distributed to customers who were told the following: Plate replacement: If the issue occurs, the surgeon should proceed as follows: 1- Unscrew the Plate screw of the cage with the screwdriver, 2- Scrap the current plate assembly, 3- Use a new set of pre-assembled plate/screw (a minimum of 2 sets of pre-assembled plates/screws per size are part of the device listing). The cages are to be used with the corresponding plate or higher plate which means that a plate with the same height than the cage can be used but also a plate with higher height. 4- In the case of 0-degree and 8-degree cages, a pre-assembled plate/screw is not mandatory as described in the surgical technique Acknowledge receipt of this letter by emailing to orders@spineart.us Discontinue use and return any product on hand. Customers with questions can contact firm at 949-313-6400 Monday through Friday, 8:00am to 5:00pm, Pacific, or email at orders@spineart.us
- Initiated
- 2023-03-16
- Root cause
- Process change control
- Distribution
- US Nationwide distribution in the states of NH, TX, CA, VA, IL, RI, NC, AZ.
- Recall number
- Z-1426-2023
- Recall status
- Open, Classified
- CDRH record ID
- 199335
- FDA classified
- 2023-04-24
- Recall event number
- 91973