Open original FDA record ↗Recall record
- FDA identifier
- Z-1426-2026
- Company in FDA record
- Centinel Spine, Inc.
- FDA classification date
- Feb 17, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Products were mislabeled as the 6mm product but included the 5 mm product.
Source facts
- Action
- On or around December 31, 2025 Phone calls were made to customers to notify them of the recall. Follow-up URGENT PRODISC C SK US IMPLANT LARGE 6MM DEVICE RECALL letters were also emailed to customers. Actions to be taken: We request that you review your inventory and return any units in your possession of this impacted lot number. We recommend monitoring any patient that has received the recalled device for any unexpected device-related symptoms throughout the regularly scheduled check-in procedures. Actions to Be Taken by Centinel Spine: Centinel Spine intends to remove all remaining implants effected by this recall from the market and issue a replacement product with corrected labeling for any remaining products in the Distributors possession that are subject to this recall.
- Initiated
- 2025-12-31
- Root cause
- Process control
- Distribution
- US Nationwide distribution in the states of AL, CA, CO, LA, NY, OR, TN, TX.
- Recall number
- Z-1426-2026
- Recall status
- Open, Classified
- CDRH record ID
- 218137
- FDA classified
- 2026-02-17
- Recall event number
- 98282