Skip to content

Public FDA record

Z-1427-2017: Juno DRF; Model: 709020 Product Usage: General, multi-function x-ray systems capable of routine radiogra

Recall record Z-1427-2017 for Villa Sistemi Medicali S.P.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1427-2017
Company in FDA record
Villa Sistemi Medicali S.P.A.
FDA record creation date
Mar 9, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position.

Source facts

Action
Customers were notified of the recall via DHL or similar mail service. The letter identified the affected product, problem and actions to be taken.
Initiated
2016-12-27
Root cause
Nonconforming Material/Component
Terminated
2021-03-09
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NJ, NY, PA, Puerto Rico, TN, TX, UT, and W.I
Recall number
Z-1427-2017
Recall status
Terminated
CDRH record ID
152981
FDA record created
2017-03-09
Recall event number
76367

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗