Z-1427-2022: GE Healthcare Avance CS2, Avance CS2 Pro Anesthesia Systems, Part #1009-9050-000; and Aisys CS2 Anesthesi

Recall record Z-1427-2022 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1427-2022
Company in FDA record
GE Healthcare, LLC
FDA record creation date
Jul 20, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The base of the anesthesia system can have a crack and could potentially fracture resulting in a tip or overbalance of the anesthesia system. This can result in potential injury if it falls on a person.

Source facts

Action
The recalling firm issued letters dated 5/27/2022 via traceable means on 5/27/2022 which were addressed to the Director of Respiratory, Chief of Anesthesia, Health Care Administrator/Risk Manager, and Director of Biomedical/Clinical Engineering. The letter is flagged that the consignee needs to ensure all potential users in the facility are made aware of this safety notification and recommended actions. The notification informs the consignees of the affected devices, the safety issue, and actions to be taken. The actions notified the consignee they can continue to use the anesthesia system by following instructions listed in the letter. The instructions were to inspect the anesthesia system for cracks at the rear of the base as depicted in the diagram and photograph, and if no cracks are observed, they can continue to use the anesthesia system. If a crack at the rear is observed, they can continue to use the system under the following conditions: Limit the movement of the system using care on uneven flooring/threshold and ensure the floor is clear of obstacles (cables, power cords, etc.); and do not exceed the recommended weight limit for equipment mounted or supported by the anesthesia system. A Medical Device Notification Acknowledgement Response form was to be completed and returned within 30 days of receipt to acknowledge receipt and understanding of the notification. The letter informs the consignee that the recalling firm will inspect all affected devices at no cost and that a firm representative will contact the consignee to arrange for the correction.
Initiated
2022-05-27
Root cause
Process control
Distribution
Distribution was made to AZ, CA, CO, CT, FL, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MO, MT, NC, NE, NV, NY, OH, PA, SC, TN, TX, VA, and WV. There was also military distribution but no government distribution. Foreign distribution was made to Australia, Canada, China, Estonia, Finland, France, Germany, Greece, India, Indonesia, Iraq, Ireland, Italy, Japan, Jordan, Kenya, Korea, Malaysia, New Zealand, Philippines, Romania, Russia, Saudi Arabia, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
Recall number
Z-1427-2022
Recall status
Open, Classified
CDRH record ID
194360
FDA record created
2022-07-20
Recall event number
90422

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗