Z-1427-2023: Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set, Reference Part Numbers C-PTIS-100-HC-G-EU-F

Recall record Z-1427-2023 for Cook Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1427-2023
Company in FDA record
Cook Incorporated
FDA classification date
May 10, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

There were reports from customers that the tracheostomy in the sets and trays is not making a secure connection with the 15mm cap and other 15mm circuit components and accessories. This resulted in an unsecure connection between the device connector and circuit components, cap, or accessories.

Source facts

Action
An URGENT: MEDICAL DEVICE RECALL dated 4/7/23 was sent to customers. Actions to be Taken by the Customer 1. Examine inventory immediately to determine if you have Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Sets and Trays from the Cook Medical lots in Attachment A. If so, quarantine any affected sets/trays that remain unused. Immediately cease all further distribution and use of the affected sets/trays. 2. Return the complete sets/trays to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected sets/trays on hand. You may return the Form via fax (812.339.7316) or email (FieldActionsNA@CookMedical.com). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected sets/trays have been transferred. 5. Please immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: CustomerRelationsNA@CookMedical.com. Action being Taken by Cook Medical Cook Medical is removing potentially affected products from the market and replacing potentially defective tracheostomy tubes with new tracheostomy tubes as they become available from Medtronic. We recognize that this situation is a potential disruption to your normal operations, and we sincerely apologize. Thank you for your immediate assistance in this matter. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235. We look forward to your return of the Acknowledgement and Receipt Form.
Initiated
2023-04-07
Root cause
Nonconforming Material/Component
Distribution
Worldwide distribution - US Nationwide.
Recall number
Z-1427-2023
Recall status
Open, Classified
CDRH record ID
199620
FDA classified
2023-05-10
Recall event number
92083

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗