Open original FDA record ↗Recall record
- FDA identifier
- Z-1428-2010
- Company in FDA record
- Elekta, Inc.
- FDA classification date
- Apr 22, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Images from a different patient can be imported without a warning for the user.
Source facts
- Action
- Phase I - A technical note for LGP and SPS describing the risk for importing wrong image files and including recommendations on how to reduce this risk has been written and distributed to users of both Leksell GammaPlan and Leksell SurgiPlan on May 23, 2005 via FCO DOC 05-0001. Phase 2 - A maintenance release was created of the Scanner Interface (SIF) to correct the problem. This was distributed to Leksell SurgiPlan users on November 30, 2005 via FCO0000153-1. Additional questions are directed to the firm to Product Manager lrobert.aslund AT elekta.com, or After Sales gustaf.piehl AT elekta.com..
- Initiated
- 2005-05-23
- Root cause
- Other
- Terminated
- 2010-12-14
- Distribution
- Nationwide, Canada and Mexico.
- Recall number
- Z-1428-2010
- Recall status
- Terminated
- CDRH record ID
- 90064
- FDA classified
- 2010-04-22
- Recall event number
- 55144