Z-1428-2020: HeartWare HVAD System Battery Charger AC Adapter The HeartWare HVAD System is indicated for hemodynamic

Recall record Z-1428-2020 for Heartware, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1428-2020
Company in FDA record
Heartware, Inc.
FDA classification date
Mar 9, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Potential for an HVAD System user to mistakenly insert the Battery Charger AC Adapter into an HVAD Controller power port.

Source facts

Action
Heartware notified customers on about 01/23/2020 via "Urgent Field Safety Notice" letter. Customers were notified of the potential for an HVAD System user to mistakenly insert the Battery Charger AC Adapter into an HVAD Controller power port. When this occurs, communication circuits between the battery and the Controller may be damaged. If circuit damage occurs, it will be necessary to exchange the Controller to restore communication between the batteries and Controller. Instructions included to advise patients to carefully follow the guidance provided in the patient manual related to connecting a power source. Pay attention when connecting and disconnecting power supply cables to ensure users verify the correct cable is being connected to their Controller. Patients were instructed to report any persistent, unexpected audible tones to their Clinician for additional instructions. Additional instructions included to ensure all personnel within your organization or any customers if further distributed are notified of this recall. If you have any questions, please contact the Cardiac and Vascular Group Customer Focused Quality at 763-526-2513 or by email to: carlos.j.alfonzo@medtronic.com.
Initiated
2020-01-23
Root cause
Device Design
Distribution
No affected devices distributed within the US. International distribution to Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, EGYPT, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Kazakhstan, Korea, Korea, Republic of, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Macedonia, Malaysia, Netherlands, New Zealand, Norway, Poland, Romania, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Taiwan, Province of China, Turkey, United Arab Emirates, United Kingdom, Vietnam.
Recall number
Z-1428-2020
Recall status
Open, Classified
CDRH record ID
179518
FDA classified
2020-03-09
Recall event number
84829

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗