Open original FDA record ↗Recall record
- FDA identifier
- Z-1428-2022
- Company in FDA record
- Aesculap Implant Systems LLC
- FDA record creation date
- Jul 21, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Fracture or breakage of the spinal fixation arm at the downtube instrument if the 90 degree angle is not met when using the removal key with the spinal fixation system.
Source facts
- Action
- An Urgent Field Safety Notification was issued July 13, 2020 sent USPS certified mail to distributors. The notification identified the product as Ennovate Removal Key (SZ380R) and addressed correct usage of said device, referencing section K of the Surgical Technique. Distributors are to identify and notify customers if further distributed. Distributors and customers are to review the notice, notify end users, and return the "Medical Device Field Safety Notification Form" to Aesculap via email at qa-recalls@aesculapusa.com.
- Initiated
- 2020-07-13
- Root cause
- Component design/selection
- Terminated
- 2024-04-26
- Distribution
- United States Nationwide distribution in the states of CA, CO, IL, MI, OH & TX.
- Recall number
- Z-1428-2022
- Recall status
- Terminated
- CDRH record ID
- 194602
- FDA record created
- 2022-07-21
- Recall event number
- 90503