- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Due to a consumer complaint, North Coast Medical determined that due to a design flaw, there was not enough plastic support material to retain bracket bolts when subjected to a force greater than 150 lbs.
Source facts
- Action
- North Coast Medical, Inc. sent a Recall Notification Letter dated March 20, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to identify and discard the affected product. Complete the enclosed Form and return the form even if they no longer have the bath bench. Return the form by fax to 408-775-1430. Customers should contact the Quality Department if they have any questions, concerns or need any assistance following the instructions. For questions regarding this recall call 800-821-9319, ext 121.
- Initiated
- 2012-03-20
- Root cause
- Device Design
- Terminated
- 2012-07-20
- Distribution
- Nationwide Distribution
- Recall number
- Z-1429-2012
- Recall status
- Terminated
- CDRH record ID
- 108141
- FDA classified
- 2012-04-16