Z-1429-2020: Seca 402/403 Baby Scale Cart, Ref # 4020000009/ REF # 4030000009

Recall record Z-1429-2020 for Seca. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1429-2020
Company in FDA record
Seca
FDA record creation date
Mar 3, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It has been determined that the wheels on the baby scale cart can unexpectedly loosen and in the worst case, detach from the cart completely. If this occurs the cart can become unstable and tip over, potentially leading to the scale falling from the cart. If a baby is left unattended in a cart, the baby could fall off the scale and be subjected to several serious injuries.

Source facts

Action
On June 19, 2019 a "Urgent: Correction/recall of seca 402/403 - Cart for seca baby scale" was mailed to customers. The letter provided details on the recall product and asked customers to do the following: End Users: Immediate Measures to Take: " With the appropriate Allen wrench, re-tighten all the screw connections as firmly as possible (see attachment 1). " Apply the brakes on both wheels. " Mark the baby scale cart with the warning label (see attachment 2) " Do not make mobile use of the cart, use it only at one fixed location. " Please confirm receipt of this notice by completing attachment 3, scanning the attachment, and returning it to seca by e-mail to babyscalecart.us@seca.com within one week of receipt of this information " Wait for further information about Upgrade Kit to replace the affected wheels (see section Upgrade Kits). DISTRIBUTORS: Immediate Measures to Take " Please stop the delivery of the baby scale carts right away. " If you have any carts in stock, quarantine the carts by placing them in one specific location. " Wait for further information about Upgrade Kit to replace the affected wheels. (see section Upgrade Kits). " Please IMMEDIATELY forward a of the Urgent Correction/Recall of seca 402/403 for end users to your end users (attached to this letter). seca will be reaching out to all distributors and end users directly, out of overabundance of caution. Therefore, please identify the customers who have purchased seca 402/403 and send us the customer list within one week of receipt of this information. The list shall include name and address of the customers, number of the purchased carts and their serial numbers, names of contact persons, and their email addresses. " Please confirm receipt of this Correction/Recall notice by scanning and returning to us the response form in attachment 2 by e-mail to babyscalecart.us@seca.com within one week of receipt of this information. Upgrade Kits: Seca is already working
Initiated
2019-06-19
Root cause
Process design
Distribution
US: CA, MI, WA, WV, VA, NM, LA, TN, IL, PA, OH, FL, WA, OR, NY, CO, IA, MN, TX, MA, NC, LA, KY, NM AL,DC, NJ, AK, AZ, NV, SC, HI, MD, WI, RI, DE, MO OUS: Canada
Recall number
Z-1429-2020
Recall status
Open, Classified
CDRH record ID
179463
FDA record created
2020-03-03
Recall event number
84817

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗