Open original FDA record ↗Recall record
- FDA identifier
- Z-1429-2022
- Company in FDA record
- Aesculap Implant Systems LLC
- FDA record creation date
- Jul 21, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Fracture or breakage of the spinal fixation arm at the downtube instrument if the 90 degree angle is not met when using the removal key with the spinal fixation system. Redesign of removal key.
Source facts
- Action
- An Urgent Field Safety Notification was issued November 18, 2020 sent USPS certified mail to distributors regarding removal of Ennovate Removal Key Short (SZ380) from the field. The affected Ennovate MIS Removal Key Short (SZ380R) in the field is to be exchanged with an improved design version under the same article number manufactured after June 01, 2020. Distributors are to identify and notify customers of this action. Distributors and customers are to identify the affected version in inventory and return it to Aesculap and submit the Acknowledgment Form via email at qa-recalls@aesculapusa.com.
- Initiated
- 2020-11-18
- Root cause
- Component design/selection
- Terminated
- 2024-12-23
- Distribution
- United States Nationwide distribution in the states of CA, CO, MI, NH, OH, SC & TX.
- Recall number
- Z-1429-2022
- Recall status
- Terminated
- CDRH record ID
- 194641
- FDA record created
- 2022-07-21
- Recall event number
- 90526