Z-1429-2023: HAMILTON-C6, REF: 160021

Recall record Z-1429-2023 for Hamilton Medical AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1429-2023
Company in FDA record
Hamilton Medical AG
FDA classification date
Apr 25, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software error causes, safety ventilation, in which ventilation continues in the "safety ventilation" mode with audible/visible alarm - patient inputs are not monitored, if the following coincide 1) A mode change to an adaptive mode (ASV, APVcmv, APVsimv, INTELLiVENT-ASV, (S)CMV+, SIMV+), and 2) The controller and/or humidifier is connected to the ventilator and is operational.

Source facts

Action
On 3/14/23, correction notices were emailed to health care facilities and distribution partners. Customers were told the following: 1) Firm will reach out to schedule the software update. 2) See Operation's Manual, Section 7.7, Safety Ventilation. The device must turn off to exit Safety Ventilation. 3) Complete and return the confirmation form to bret.everett@hamiltonmedical.com 4) If the malfunction occurs take the following actions: a) Monitor your patient. b) Provide an alternative device for ventilation. c) Have the ventilator serviced. If you have questions or concerns, contact Technical Support at 1-800-426-6331, option #2, or email Reno.TechSupport@hamiltonmedical.com
Initiated
2023-03-14
Root cause
Software design
Distribution
US Nationwide distribution in the states of CA, MA, ID, NC, GA, FL, NV, PA, MO, KS, SC, NY, NM, WY, VA.
Recall number
Z-1429-2023
Recall status
Open, Classified
CDRH record ID
199226
FDA classified
2023-04-25
Recall event number
91925

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗