Open original FDA record ↗Recall record
- FDA identifier
- Z-1429-2023
- Company in FDA record
- Hamilton Medical AG
- FDA classification date
- Apr 25, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software error causes, safety ventilation, in which ventilation continues in the "safety ventilation" mode with audible/visible alarm - patient inputs are not monitored, if the following coincide 1) A mode change to an adaptive mode (ASV, APVcmv, APVsimv, INTELLiVENT-ASV, (S)CMV+, SIMV+), and 2) The controller and/or humidifier is connected to the ventilator and is operational.
Source facts
- Action
- On 3/14/23, correction notices were emailed to health care facilities and distribution partners. Customers were told the following: 1) Firm will reach out to schedule the software update. 2) See Operation's Manual, Section 7.7, Safety Ventilation. The device must turn off to exit Safety Ventilation. 3) Complete and return the confirmation form to bret.everett@hamiltonmedical.com 4) If the malfunction occurs take the following actions: a) Monitor your patient. b) Provide an alternative device for ventilation. c) Have the ventilator serviced. If you have questions or concerns, contact Technical Support at 1-800-426-6331, option #2, or email Reno.TechSupport@hamiltonmedical.com
- Initiated
- 2023-03-14
- Root cause
- Software design
- Distribution
- US Nationwide distribution in the states of CA, MA, ID, NC, GA, FL, NV, PA, MO, KS, SC, NY, NM, WY, VA.
- Recall number
- Z-1429-2023
- Recall status
- Open, Classified
- CDRH record ID
- 199226
- FDA classified
- 2023-04-25
- Recall event number
- 91925