Software anomaly in the patient positioning system may result in positional discrepancy.
Source facts
Action
Between the dates of February 5-11, 2026, Hitachi, Ltd, issued an "URGENT: MEDICAL DEVICE RECALL" notification dated January 21, 2026, via Service Team Manager to affected consignees. Hitachi is asking consignees to take the following actions:
1. Until the PPS software has been updated with the corrected version, users are instructed to follow the System Operation instruction in the document.
2. Users are also asked to post the document in a prominent space and sign its cover.
3. Complete and return the acknowledgment and receipt form.
Action by the firm: Hitachi plans to update the PPS software to a corrected version that prevents the mismatch between the commanded error correction parameter and the internal variable. starting in late March 2026 and the users will be contacted after the update.
If you have any questions, please call QA Manager at Tel: (408) 986-6300. Monday through Friday, 8:00 AM to 5:00 PM, Central Time.
Initiated
2026-02-06
Root cause
Software design
Distribution
US distribution to states of: TX, Washington DC
Recall number
Z-1429-2026
Recall status
Open, Classified
CDRH record ID
218304
FDA classified
2026-02-20
Recall event number
98330