Z-1429-2026: Brand Name: Hitachi Proton Beam Therapy System Ref: PROBEAT-FR

Recall record Z-1429-2026 for Hitachi, Ltd. Radiation Oncology Systems, Kashiwanoha. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1429-2026
Company in FDA record
Hitachi, Ltd. Radiation Oncology Systems, Kashiwanoha
FDA classification date
Feb 20, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software anomaly in the patient positioning system may result in positional discrepancy.

Source facts

Action
Between the dates of February 5-11, 2026, Hitachi, Ltd, issued an "URGENT: MEDICAL DEVICE RECALL" notification dated January 21, 2026, via Service Team Manager to affected consignees. Hitachi is asking consignees to take the following actions: 1. Until the PPS software has been updated with the corrected version, users are instructed to follow the System Operation instruction in the document. 2. Users are also asked to post the document in a prominent space and sign its cover. 3. Complete and return the acknowledgment and receipt form. Action by the firm: Hitachi plans to update the PPS software to a corrected version that prevents the mismatch between the commanded error correction parameter and the internal variable. starting in late March 2026 and the users will be contacted after the update. If you have any questions, please call QA Manager at Tel: (408) 986-6300. Monday through Friday, 8:00 AM to 5:00 PM, Central Time.
Initiated
2026-02-06
Root cause
Software design
Distribution
US distribution to states of: TX, Washington DC
Recall number
Z-1429-2026
Recall status
Open, Classified
CDRH record ID
218304
FDA classified
2026-02-20
Recall event number
98330

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗