Skip to content

Public FDA record

Z-1430-2020: Remisol Advance running on hardware with Windows XP, Windows 7, Windows Server 2003 and Windows Server 20

Recall record Z-1430-2020 for Normand-Info S.A.S.U.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1430-2020
Company in FDA record
Normand-Info S.A.S.U.
FDA record creation date
Mar 3, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A patch was released for a critical remote code execution vulnerability in remote desktop services. This vulnerability can be exploited remotely without authentication on certain operating systems. An attacker who successfully exploited this vulnerability could execute arbitrary code on the target system; then install programs; view, change, or delete data; or create new accounts with full user rights. If the vulnerability could be exploited locally, it would lock down the computer.

Source facts

Action
On 08/15/19, Urgent Medical Device Recall notices, dated 08/09/19, were mailed to customers informing them to follow the instructions in the notice to download the patch for their system(s). In addition, customers were asked to share this information with laboratory staff, to retain the notification as part of their laboratory Quality System documentation, and to complete and return the response form. If you are a centralized license holder, please provide the other affected laboratories of your organization or association with a copy of this letter. The latest information on cybersecurity can be found on the following website: https://www.beckmancoulter.com/en/about-beckman-coulter/product-security/product-security-updates. Customers with additional questions were encouraged to contact the firm via email at website: http://www.beckmancoulter.com. The following number was also provided: (800) 526-3821
Initiated
2019-08-15
Root cause
Other
Terminated
2021-06-03
Distribution
U.S.: NY, AZ, FL, NJ, MA, OH, CA, IL, NC, OK, IN, KY, OR, MI, WI, PA, UT, GA, VA, TX, NE, SD, IA, MN, MD, DE, SC, MT, WV, LA, MO, TN, NM, NV, HI, WA, WY, AL, AR, MS, RI, ID, ME, DC, ND, AK, CO, CT, KS, NH, VT
Recall number
Z-1430-2020
Recall status
Terminated
CDRH record ID
176140
FDA record created
2020-03-03
Recall event number
83687

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗