- FDA identifier
- Z-1430-2021
- Company in FDA record
- Boston Scientific Corporation
- FDA record creation date
- Apr 15, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm initiated a removal of certain batches of MAMBA 135 and MAMBA Flex 135 Microcatheters due to incorrect labeling and packaging.
Source facts
- Action
- A Recall notification letter titled, "Medical Device Removal - Immediate Action Required" was sent to the OUS consignees on 03/05/2021 and the U.S. consignees on 03/08/2021 via overnight mail delivery. Customers outside of the US were either contacted via letter delivery or email. The recalling firm requested if the consignees identify any product from the listed batch within their inventory, to please segregate the product immediately and return it to Boston Scientific in accordance with the enclosed instructions. The consignees will receive replacement product for all removed product returned to Boston Scientific. Distributors are also informed in the letter that the removal depth is to the hospital level and the removal notification should be forwarded to their customers. The field removal letter provides customers with the product segregation and return instructions and accounts are asked to complete a Reply Verification Tracking Form indicating they have received the field removal and followed the included instructions.
- Initiated
- 2021-03-05
- Root cause
- Labeling False and Misleading
- Terminated
- 2023-01-05
- Distribution
- Distributed nationwide to AL, AR, AZ, CO, FL, GA, IA, IL, KY, LA, MI, MN, MO, MS, NC, ND, NV, NY, OK, OH, PR, SC, TN, VA, WA, WI and internationally to Belgium, France, Great Britain, Israel, Italy, Lithuania, Netherlands, Poland, Puerto Rico, Switzerland
- Recall number
- Z-1430-2021
- Recall status
- Terminated
- CDRH record ID
- 186251
- FDA record created
- 2021-04-15
- Recall event number
- 87539