Skip to content

Public FDA record

Z-1430-2021: MAMBA Microcatheter. Packaged with protective sleeve over catheter, sealed in a Tyvek pouch, and is then

Recall record Z-1430-2021 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1430-2021
Company in FDA record
Boston Scientific Corporation
FDA record creation date
Apr 15, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm initiated a removal of certain batches of MAMBA 135 and MAMBA Flex 135 Microcatheters due to incorrect labeling and packaging.

Source facts

Action
A Recall notification letter titled, "Medical Device Removal - Immediate Action Required" was sent to the OUS consignees on 03/05/2021 and the U.S. consignees on 03/08/2021 via overnight mail delivery. Customers outside of the US were either contacted via letter delivery or email. The recalling firm requested if the consignees identify any product from the listed batch within their inventory, to please segregate the product immediately and return it to Boston Scientific in accordance with the enclosed instructions. The consignees will receive replacement product for all removed product returned to Boston Scientific. Distributors are also informed in the letter that the removal depth is to the hospital level and the removal notification should be forwarded to their customers. The field removal letter provides customers with the product segregation and return instructions and accounts are asked to complete a Reply Verification Tracking Form indicating they have received the field removal and followed the included instructions.
Initiated
2021-03-05
Root cause
Labeling False and Misleading
Terminated
2023-01-05
Distribution
Distributed nationwide to AL, AR, AZ, CO, FL, GA, IA, IL, KY, LA, MI, MN, MO, MS, NC, ND, NV, NY, OK, OH, PR, SC, TN, VA, WA, WI and internationally to Belgium, France, Great Britain, Israel, Italy, Lithuania, Netherlands, Poland, Puerto Rico, Switzerland
Recall number
Z-1430-2021
Recall status
Terminated
CDRH record ID
186251
FDA record created
2021-04-15
Recall event number
87539

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗