Skip to content

Public FDA record

Z-1431-2026: Medline Sterile Saline Wound Wash, Cleansing & Wound Irrigation, USP Normal Saline (0.9% Sodium Chloride)

Recall record Z-1431-2026 for Medline Industries, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1431-2026
Company in FDA record
Medline Industries, LP
FDA classification date
Feb 20, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Medline has identified that the manufacturer of the sterile saline wound wash may not have manufactured products to meet the minimum required sterility assurance level required for sterile products.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification letter dated 1/6/26 was sent to customers regarding Saline Wound Wash. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. Website link: https://recalls.medline.com Recall Reference #: R-25-267 Recall Code: 3. Your account will receive credit once the response form is submitted. 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers destroy any affected product. You should include your customers quantities on your response form. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com. An URGENT MEDICAL DEVICE RECALL notification letter dated 1/5/26 was sent to customers regarding Meijer Saline Wound Wash item number 347792. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed on the enclosed response form. Quarantine all affected product. Upon completion of the below actions, please destroy affected product. 2. Please complete and return the enclosed response form listing the quantity of affected product you have in inventory. Even if you do not have any affected product in inventory, please complete and return the response form. 3. Your account will receive credit once the response form is submitted, if applicable. 4. If you are a distributor, or have resold or transferred this prod
Initiated
2026-01-05
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Bahamas, Bermuda, Cayman Islands.
Recall number
Z-1431-2026
Recall status
Open, Classified
CDRH record ID
218236
FDA classified
2026-02-20
Recall event number
98319

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗