Open original FDA record ↗Recall record
- FDA identifier
- Z-1432-2010
- Company in FDA record
- Elekta, Inc.
- FDA classification date
- Apr 23, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
In some instances, the bolts which are supplied with the table have sheared off, resulting in inability to operate the powered table.
Source facts
- Action
- Mandatory Field Change Order (FCO) 528 053 "Check for zinc plated Precise Table 'Z' mechanism fixation bolts", dated 1/11/2007 was distributed to all affected customers. The Change Order identified the product, stated the reason for the FCO, and the procedure to check and resolve the problem identified.
- Initiated
- 2007-01-11
- Root cause
- Other
- Terminated
- 2010-12-28
- Distribution
- Worldwide Distribution -- USA, Puerto Rico, Argentina, Australia, Austria, Bangladesh, Belgium, Canada, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, Pakistan, Peru, Poland, Russia, Serbia, Spain, Sudan, Sweden, Taiwan, Thailand, Turkey & United Kingdom.
- Recall number
- Z-1432-2010
- Recall status
- Terminated
- CDRH record ID
- 89897
- FDA classified
- 2010-04-23
- Recall event number
- 55103