Skip to content

Public FDA record

Z-1432-2025: Orthofix Pillar SA Ti Spacer System Intervertebral Body Fusion Spinal Device, sterile, implant grade tita

Recall record Z-1432-2025 for Orthofix U.S. LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1432-2025
Company in FDA record
Orthofix U.S. LLC
FDA classification date
Mar 21, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The product is mislabeled with the incorrect anterior height of 10mm, but the laser marking on the implant and corresponding trial both show the correct anterior height of 10.5mm.

Source facts

Action
On 2/20/2025, correction notices were emailed to distributors who were informed of the following: This notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where potentially affected devices have been transferred. Attach this notice to the inventory in your possession, if possible. Should you have any questions regarding this matter, please contact the firm at madison.ebel@seaspine.com
Initiated
2025-02-20
Root cause
Labeling Change Control
Distribution
US Nationwide distribution in the states of California, Colorado, and Missouri.
Recall number
Z-1432-2025
Recall status
Open, Classified
CDRH record ID
212611
FDA classified
2025-03-21
Recall event number
96351

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.