Open original FDA record ↗Recall record
- FDA identifier
- Z-1434-2021
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Apr 19, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to the configuration of certain Windows Service Permissions within the operating systems of the Sensis/ Sensis Vibe computer, there is a risk for exposure of sensitive information, manipulation of data, or Denial of Service attacks and could result in incorrect diagnostic or therapeutic decisions
Source facts
- Action
- Siemens Healthineers issued Customer Safety Advisory Notice dated 3/4/21 to affected customers via AX074/20/S. Letter states reason for recall, health risk and action to take: Siemens will restrict the Windows Service Permissions to the required level only and the IT security vulnerability of your system will be reduced via Update Instruction AX073/20/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein. .
- Initiated
- 2021-03-04
- Root cause
- Software design
- Terminated
- 2024-09-30
- Distribution
- US Nationwide distribution.
- Recall number
- Z-1434-2021
- Recall status
- Terminated
- CDRH record ID
- 186215
- FDA classified
- 2021-04-19
- Recall event number
- 87515