Rotation of the instrument reagent compartment can cause a bubble to form in the neck of the reagent bottle causing premature liquid level detection and short sampling of the reagent, resulting in falsely low or zero patient results.
Source facts
Action
Consignees were mailed recall notices dated 8/20/04 and were instructed to either reduce the initial reagent bottle volume prior to use, or to request replacement product.
Initiated
2004-08-20
Root cause
Other
Terminated
2004-11-05
Distribution
Nationwide.
Recall number
Z-1435-04
Recall status
Terminated
CDRH record ID
34738
FDA classified
2004-09-02
Recall event number
29893