Open original FDA record ↗Recall record
- FDA identifier
- Z-1435-2011
- Company in FDA record
- AGFA Corp.
- FDA classification date
- Feb 25, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
User could modify a measurement's units and that change would not be reflected in echocardiography formulas used within the Adult echocardiography RM.
Source facts
- Action
- The firm, Agfa Healthcare, sent individual letters dating from April 15, 2009 to August 26, 2009, to the customers. The letters described the product, problem and actions to be taken. Agfa provided the affected customer with a complete list of their impacted patients and the measurements. Agfa also scheduled unit corrections with their customers and received acknowledgement of the letters at the time of scheduling the unit corrections. Service Bulletin DIS026.09E was issued January 2009 to inform users that Agfa will no longer allow customers to request unit changes for measurements that have formulas. Note: Technical Support ran diagnostic scripts for each customer and documented in GSC cases for each consignee/customer. If you have any questions, contact Agfa Healthcare at (949) 830-4959.
- Initiated
- 2009-04-15
- Root cause
- Other
- Terminated
- 2011-10-13
- Distribution
- Nationwide distribution: USA including states of: IL, MT, NY, OR and SC.
- Recall number
- Z-1435-2011
- Recall status
- Terminated
- CDRH record ID
- 97505
- FDA classified
- 2011-02-25
- Recall event number
- 57820