Skip to content

Public FDA record

Z-1435-2011: IMPAX CV Results Manager/Results Manager Administration Tool The IMPAX Cardiovascular suite is a cardiov

Recall record Z-1435-2011 for AGFA Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1435-2011
Company in FDA record
AGFA Corp.
FDA classification date
Feb 25, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

User could modify a measurement's units and that change would not be reflected in echocardiography formulas used within the Adult echocardiography RM.

Source facts

Action
The firm, Agfa Healthcare, sent individual letters dating from April 15, 2009 to August 26, 2009, to the customers. The letters described the product, problem and actions to be taken. Agfa provided the affected customer with a complete list of their impacted patients and the measurements. Agfa also scheduled unit corrections with their customers and received acknowledgement of the letters at the time of scheduling the unit corrections. Service Bulletin DIS026.09E was issued January 2009 to inform users that Agfa will no longer allow customers to request unit changes for measurements that have formulas. Note: Technical Support ran diagnostic scripts for each customer and documented in GSC cases for each consignee/customer. If you have any questions, contact Agfa Healthcare at (949) 830-4959.
Initiated
2009-04-15
Root cause
Other
Terminated
2011-10-13
Distribution
Nationwide distribution: USA including states of: IL, MT, NY, OR and SC.
Recall number
Z-1435-2011
Recall status
Terminated
CDRH record ID
97505
FDA classified
2011-02-25
Recall event number
57820

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗