Z-1435-2020: Atellica IM Humidity Pack (Qty 5), SMN 11313505, UDI 00630414234526, Software Version V1.21.0 and lower -

Recall record Z-1435-2020 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1435-2020
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA record creation date
Mar 4, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A software error is causing the analyzer to incorrectly eject affected Humidity Packs as expired.

Source facts

Action
On January 7, 2020, the firm issued an Urgent Medical Device Correction letter to all affected US customers. Customers were informed that the analyzer may erroneously identify that humidity packs Lot 0100 have expired and eject them. The firm is investigating the root cause and customers will be notified when additional information is available. The behavior will be corrected in a new version of software that will be available shortly. It is important to note that Humidity Packs do have an onboard stability (OBS) of 180 days that is monitored by the system software. Once loaded onto an analyzer, the OBS of each pack is monitored. The firm is not recommending a review of previously generated results, since the accuracy of results is not affected by this issue. Customers were given workaround instructions so that they may manually enter the humidity packs.
Initiated
2020-01-06
Root cause
Software in the Use Environment
Terminated
2022-07-11
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Recall number
Z-1435-2020
Recall status
Terminated
CDRH record ID
179477
FDA record created
2020-03-04
Recall event number
84823

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗