Reason stated in the source
A software error is causing the analyzer to incorrectly eject affected Humidity Packs as expired.
Source facts
- Action
- On January 7, 2020, the firm issued an Urgent Medical Device Correction letter to all affected US customers. Customers were informed that the analyzer may erroneously identify that humidity packs Lot 0100 have expired and eject them. The firm is investigating the root cause and customers will be notified when additional information is available. The behavior will be corrected in a new version of software that will be available shortly. It is important to note that Humidity Packs do have an onboard stability (OBS) of 180 days that is monitored by the system software. Once loaded onto an analyzer, the OBS of each pack is monitored. The firm is not recommending a review of previously generated results, since the accuracy of results is not affected by this issue. Customers were given workaround instructions so that they may manually enter the humidity packs.
- Initiated
- 2020-01-06
- Root cause
- Software in the Use Environment
- Terminated
- 2022-07-11
- Distribution
- Domestic distribution nationwide. Foreign distribution worldwide.