Z-1436-2026: MRI LED Mobile Exam Light Battery REF SL-111 This non-magnetic UL and CE approved battery is used in con

Recall record Z-1436-2026 for MRIMed Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1436-2026
Company in FDA record
MRIMed Inc.
FDA classification date
Feb 20, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to battery component overheating while charging resulting in melting of internal components and causing smoke.

Source facts

Action
On or about 10/13/2025, the firm called customers and emailed an "URGENT MEDICAL DEVICE QUARANTINE" Letter to customer and distributers informing them that the firm has become aware of the possibility of the battery overheating during charging, resulting in melting of the internal components. Customers and Distributors are instructed to: 1. Immediately discontinue use of the battery. 2. Do not charge the battery. 3. Remove the charger from the battery. 4. Label the battery and the light as BATTERY IN QUARANTINE, DO NOT USE 5. Keep the batteries in a secure location to avoid unintended use, and await further instructions from MRI Med. For Distributors are instructed to cease distribution of the affected products. For Questions - contact MRI Med, Inc. Monday through Friday, 8:00am-3:00pm, PST, or email sales@MRImed.com or via phone at 707-762-8600
Initiated
2025-10-13
Root cause
Under Investigation by firm
Distribution
Worldwide - U.S. Nationwide distribution in the states of AL, CA, PA, SC, GA, MA, TX, OH, NY, NC, IA, WA, TN, NJ, MO, LA, MN, AZ, VA, SC, KY, FL, MD, IN, AK, WI, NM, DC, RI, CT, CO, OK, DE, NE, MD, IL, NH and the countries of Germany, Canada.
Recall number
Z-1436-2026
Recall status
Open, Classified
CDRH record ID
217668
FDA classified
2026-02-20
Recall event number
98127

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗