Reason stated in the source
Alarm may be silenced- Rapid switching, on and off, of the TCM4 monitor may cause the built-in acoustical alarm to remain silent when tripped.
Source facts
- Action
- The firm initiated a recall action for the suspect device with a letter dated 2/2/2005. During a site visit to the U.S. customers conducted by the firm''s Field Service reps, the reps informed the user/customers of this problem and requested that all use of the device be discontinued, pending the formulation of a more permanent solution. Beginning on 2/8/2005, a second site visit was made to the same user/customers, at which time, the affected units were upgraded with operating software version 2.12. The firm''s recall/corrective action was scheduled for completion by 3/1/2005.
- Initiated
- 2005-02-02
- Root cause
- Other
- Terminated
- 2012-05-09
- Distribution
- Nationwide
- Recall number
- Z-1437-06
- Recall status
- Terminated
- CDRH record ID
- 46640
- FDA classified
- 2006-08-30
- Recall event number
- 35700