Skip to content

Public FDA record

Z-1437-2021: Tritanium X TL Instruments - intervertebral fusion device with bone graft, lumbar. Catalogs:TL Inserter (

Recall record Z-1437-2021 for Stryker Spine. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1437-2021
Company in FDA record
Stryker Spine
FDA record creation date
Apr 20, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Stryker identified non-conforming instruments that are components of the Tritanium X TL sets. The firm found non-conformances with the TL Inserter, TL Draw Rod, Tube Set/Syringe Adapter Assembly, and Syringe Plunger that could lead to device failure during a procedure.

Source facts

Action
On 03/17/2021, Stryker issued an Urgent Medical Device Recall notice to customers and distributers via letter because the firm identified non-conforming components of the Titanium X TL sets which could lead to device failure in the field.
Initiated
2021-03-11
Root cause
Device Design
Terminated
2023-01-05
Distribution
US Nationwide distribution in the states of AZ, CA, CO, GA, KY, MA, MT, NC, NE, NJ, NY, OH, PA, WY.
Recall number
Z-1437-2021
Recall status
Terminated
CDRH record ID
186336
FDA record created
2021-04-20
Recall event number
87580

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗