Z-1437-2023: Sterile surgical convenience kits: 1. regard Item Number: 800756005, NU00961E - OH Laminectomy Cust

Recall record Z-1437-2023 for ROi CPS LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1437-2023
Company in FDA record
ROi CPS LLC
FDA classification date
Apr 25, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The recalling firm was notified by their supplier that the light handle covers may separate from the light handle and fall off during use. The light handle covers are a kit component.

Source facts

Action
The recalling firm issued letters dated 3/22/2023 via email the same day. The letter explained the issue, identified the affected product, and listed the actions to be taken by the customer. The actions included checking their inventory for the affected product, segregate and quarantine all affected kits in their inventory, and use the attached label template to print warning labels to be affixed to all affected kits in their inventory. The label should be applied to each affected kit and outer product case in a prominent location to end users of the kits. The location chosen should not cover critical product information found on the existing label. The label instructs users to remove and discard the STERIS Light Handle Cover(s) when the kit is opened for use and to pull a sterile replacement from stock. A reply form is included with the letter for completion to indicate the consignee's actions in response to the notice. The response form is to be returned to the recalling firm via email. If the direct account consignee further distributed or forwarded the affected product, they are to immediately notify those customers. A copy of the ROi CPS recall letter, the letter from STERIS, and the attached label template may be included.
Initiated
2023-03-22
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution in the states of FL and MO. There was no foreign/military/government distribution.
Recall number
Z-1437-2023
Recall status
Open, Classified
CDRH record ID
199370
FDA classified
2023-04-25
Recall event number
91984

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗